Developmental Dysplasia of the Hip
Tracking Pixel
presents
Wheeless' Textbook of Orthopaedics

Complications related to modularity of total hip components


Barrack RL. Burke DW. Cook SD. Skinner HB. Harris WH. Journal of Bone & Joint Surgery - British Volume. 75(5):688-92, 1993 Sep. We report complications from the use of modular components in 20 hip replacements in 18 patients. Fifteen complications (in 13 patients) were related to failure of a modular interface after operation. Femoral head detachment from its trunnion was seen in 6 hips from trauma (3), reduction of a dislocation (2), and normal activity (1). In one case the base of the trunnion fractured below an extra-long modular head. In seven other hips the modular polyethylene liner dislodged from its shell, causing severe damage to the shell in four cases with extensive metallosis. In one other hip an asymmetrical polyethylene liner rotated, resulting in impingement of the femoral component and recurrent dislocation. Operative errors were seen in five cases: implantation of a trial acetabular component in one; and mismatching between the size of the femoral head and the acetabular component in the others. Surgeons who use hip replacements with modular components should be aware of the potential for operative error and of the importance of early treatment for postoperative mechanical failure.



Original Text by Clifford R. Wheeless, III, MD.